Detail AA-3500 Airborne Particulate Monitor, Air-Aide Model AA-3500 Airborne Particulate Monitor, Air-Aide Airborne Particulate Monitor Model AA-3500
Airborne Particulate Monitor
Model : AA-3500
Product : Haz-Dust
Specification :
Standard System Configuration
• Air-Aide Model AA-3500 Airborne Particle Monitor
• Universal Transformers for 120 and 220/ 230 VAC.
• Includes flow audit meter and adapter
• Adjustment tool for flow rate
• Downloading and analysis software for XP
• 9-to-9-pin RS232 cable for downloading
data to a PC.
Hardware Features
• Ambient-resolution light scattering sensor.
• Liquid crystal display ( LCD) .
• Long-life brush less pump
• Unit’ s handle swivels for ergonomic use.
• High capacity NiMH
rechargeable battery.
Resolution, Stability and Operating Range
• Resolution: 0.001 mg/ m3
• Stability at constant temperature ( 2 sigma) : + 0.004 mg/ m3
• Temperature stability ( 2 sigma) : + 0.0025 mg/ m3/ oC.
• Operating Range: 0.001 to 20.0 mg/ m3 or 0.01 to 200 mg/ m3 ( Optional)
• Analog Output 0-2VDC AA-3500 Software Features
• On-screen programming of sampling and data storage/ averaging parameters.
• Computation of time-weighted average, as well as minimum
and maximum values ( date and time stamped) by location code.
• Selection of built-in factors for fine and coarse particulate matter.
• Up to 99 location codes to tag data.
• User-definable audible alarm Calibration
• Gravimetric Reference NIST Traceable – SAE Fine Test Dust
ISO 12103-1 Standard
Data Display, Storage and Averaging
• Data display rate: 1 sec ( 3-second time constant) .
• Capacity of 21, 600 data points, date and time stamped.
• User-selectable data storage/ averaging intervals:
- Short term: 2, 4 and 10 sec.
- Long term: 1, 10 and 30 min.
PC-Based DustComm Pro Software
• RS232 cable provided for convenient downloading of stored data.
• Windows-based software graphs values and computes statistics
by location code. Data can be standardized based upon filter-
based measurements. Files stored in comma-delimited,
spreadsheet-compatible format with time, date, location code
and Paint & Draw Features.
Power Requirements and Battery
• The unit operates under battery or AC power continuously
• Unit provided with universal transformers for 120 and 220/ 240
VAC.
• 10-hour standard battery lifetime during sampling ( on full
charge) .
• Power consumption at 2 l/ min: 350 mA.
• 6 VDC NiMH rechargeable battery with 3.6 A-hour storage capacity.
Dimensions and Weight
• AA-3500 Monitor: 3” H x 6” W x 9” D
• Weight: 5 lb
Tampilkan Lebih Banyak